• imc CANSAS-HCI8 misst Temperaturen und Kleinspannungen auf bis zu 800 V Niveau

    imc CANSAS-HCI8

    Measurement modules for high-isolated temperature measuring up to 800 V

HCI8

CAN Modules for high-isolated temperature measuring up to 800 V

The imc CANSAS-HCI8 measurement module allows safe and precise measurement of temperatures and low voltage signals riding on a high or indeterminate common mode level of up to 800 V. The modules are in accordance per EN 61010 device safety standards and have an isolation rating of: 800 V CAT I, 300 V CAT II.


imc CANSAS-SL-HCI8-8T with 2x4 channel LEMO inputs (waterproof per IP65)

imc CANSAS-SL-HCI8-8T-0B
imc CANSAS-SL-HCI8-8T-0B (Front)

The imc CANSAS-SL-HCI8-8T-0B module allows for safe and precise measurement of 8 thermocouples (type K) which run on high common mode voltage levels of up to 800 V. Moreover, it is designed to be especially protected (per IP65), thus being particularly well-suited for applications in e-mobility.

Measurement inputs:
LEMO thermocouple sockets, 2x4 channel, for high voltages

Isolation strength: per IEC (EN) 61010-1:2001
Isolation: 800 V CAT I, 300 V CAT II

Supported measurement mode:

  • Type-K thermocouple

Operating conditions:
Temperature range: -40°C to 85°C
Shock vibration rating: MIL STD810F
IP rating: IP65

Applications:
Temperature measurement of high-voltage components for electric and hybrid vehicles, such as batteries, fuel cells and power supply circuits.

imc CANSAS-L-HCI8-8T with 2x4 channel LEMO inputs

imc CANSAS-L-HCI8-8T-0B
imc CANSAS-L-HCI8-8T-0B (Front)

The imc CANSAS-L-HCI8-8T-0B module allows for safe and precise measurement of 8 thermocouples (type K) which run on high common mode voltage levels of up to 800 V.

Measurement inputs:
LEMO thermocouple sockets, 2x4 channel, for high voltages

Isolation strength: per IEC (EN) 61010-1:2001
Isolation: 800 V CAT I, 300 V CAT II

Supported measurement mode:

  • Type-K thermocouple

Operating conditions:
Temperature range: -40°C to 85°C
Shock vibration rating: 50g pk (5 ms)
IP rating: IP40

Applications:
Temperature measurement of high-voltage components for electric and hybrid vehicles, such as batteries, fuel cells and power supply circuits.


L-Version with PHOENIX spring-clamp terminals

imc CANSAS-L-HCI8
imc CANSAS-L-HCI8 (Front)

The imc CANSAS-L-HCI8 module enables isolated measuring of sensitive signals over 8 analog channels running on high common mode voltage levels of up to 800 V.

Measurement inputs: PHOENIX spring-clamp terminals

Isolation strength: per IEC (EN) 61010-1:2001

Supported measurement modes:

  • Low voltage
  • Resistance
  • Current (20 mA)
  • Temperature (thermocouple and Pt100)

imc CANSAS-L-HCI8-2T with 8x1 channel thermocouple sockets for high voltage

imc CANSAS-L-HCI8-2T
imc CANSAS-L-HCI8-2T (Front)

The imc CANSAS-L-HCI8-2T module enables high-isolated temperature measuring over 8 channels. The measurement inputs are type-K thermocouple sockets for high voltage.

Measurement inputs:
LEMO thermocouple sockets, 8x1 channel, for high voltages

Isolation strength:  per IEC (EN) 61010-1:2001
Isolation: 800 V CAT I, 300 V CAT II

Operating conditions:
Temperature range: -40°C to 85°C
Shock vibration rating: 50g pk (5 ms)
IP rating: IP40

Supported measurement mode:

  • Type-K thermocouple

Top